Testing Method
|
Low pressure ion exchange liquid chromatography
|
Testing Item
|
glycated hemoglobin HbA1c
|
Testing Scope
|
3% ~ 18%
|
Testing Parameters
|
Precision (CV) ≤3%, Accuracy (V) ≤1.50%
|
Testing Time
|
4 minutes (reporting), Full-processing 6 minutes (including cleaning column and recovery time)
|
Sample type
|
venous blood (EDTA anticoagulation),Peripheral finger blood
|
Sample volume
|
5μl whole blood for each sample
|
|
Functional Parameters
|
Photometer
|
415nm LED Integral flow colorimeter
|
Sampling mode
|
Full-auto 25-position sample turntable (20 sample positions, 2 quality control positions, 1 emergency position, 1 cleaning position, 1 zero position)
|
Testing mode
|
auto test for batch, single test for emergency
|
Calibration
|
Optional selection of 2 points or 4 points calibration, optional selection of manual or automation, Equipped with 2 groups of high and low level calibrators
|
Thermostatic control
|
Column and reagent thermostatic control: 25 ± 0.5℃
|
Reagent matching
|
one piece of column and reagent consumable/300 pieces of testing samples
|
Reagent allocation
|
Each standard package includes A, B, C, D eluent, calibrator, hemolytic agent, pump tubing
|
|
Output Parameters
|
Display
|
320x240 LCD Graphic Display, displaying real-time testing curve
|
Printer
|
58mm Thermal printer, printing testing curve and report
|
Report Output
|
IFCC concentration value, NGSP area percentage, ADAG average glucose
|
Data storage
|
1000 test report (including testing curve)
|
Communication interface
|
RS232 communication interface, connecting to HIS/LIS system
|
|
Working Parameters
|
Power Supply
|
AC220V±22V 50Hz 65VA
|
size
|
408mm×390mm×230mm
|
Weight
|
8Kg
|
Working Environment
|
Temperature 12℃-30℃,Relative humidity ≤85%
|